Biology, Genetics & Clinical Lab Cell Culture & Microbiology ISO 6887 decimal dilution practice

Serial Dilution Calculator

Tell this calculator how much sample you transfer at each step, how much diluent it goes into, and how many steps you plan. It returns the dilution factor per step, the cumulative dilution and the concentration in every tube of the series, plus the total diluent you need to set out. It works for ten-fold microbiology series, two-fold antibody and MIC titrations, and any other ratio you choose, and it flags the case where the volumes you entered produce no dilution at all.

Calculator

This calculator runs in your browser. Enable JavaScript for live results — the inputs, formula and worked example below remain fully readable without it.

Inputs this calculator takes, with typical values
InputWhat to enterExample
Stock concentrationThe concentration of the undiluted stock, in whatever unit you work in — cells/mL, CFU/mL, µg/mL or molar.1 ×10⁶ per mL
Sample transferred per stepThe volume you carry forward from each tube into the next one.1 mL
Diluent already in each tubeThe volume of buffer, broth or medium waiting in each tube before the transfer arrives.9 mL
Number of dilution stepsHow many transfers you make. Six ten-fold steps take a stock down by a factor of one million.6

It returns

  • Dilution factor per step — One part sample in this many parts of final volume.
  • Cumulative dilution after all steps
  • Concentration in the last tube
  • log₁₀ of the cumulative dilution
  • Total diluent to set out

The formula

DF=Vs+VdVs
Cn=C0DFn

In plain text: DF = (V_sample + V_diluent) / V_sample ; C_n = C₀ / DF^n

  • DFDilution factor of a single step (dimensionless)
  • V_sVolume of sample transferred into each tube (mL)
  • V_dVolume of diluent already in each tube (mL)
  • nNumber of transfers performed (steps)
  • C₀Concentration of the undiluted stock (per mL)
  • C_nConcentration in tube n (per mL)

The dilution factor is a ratio of sample to FINAL volume, not sample to diluent. One part into nine parts of diluent gives a final volume of ten parts and a factor of 10.

Updated Category Cell Culture & Microbiology Verified against published test cases Reading time 12 min

What a serial dilution does and why one big dilution will not do

A serial dilution reaches a very small concentration by repeating a modest dilution several times, instead of attempting one enormous one. Six ten-fold steps dilute a stock a million-fold using nothing but a 1 mL pipette and ten 10 mL tubes. Achieving the same million-fold dilution in a single step would mean measuring 1 µL into a litre — a volume you cannot deliver accurately and a vessel you cannot mix properly.

That is the whole argument for the technique. Accuracy is dominated by the smallest volume you have to measure, and a series keeps every measured volume in the range where your pipette is actually specified to perform. The trade is that errors accumulate across the steps rather than occurring once, which is why the number of steps should be the smallest that reaches your target.

The technique is the backbone of viable counting, where you need plates in the countable window and cannot know the titre in advance; of antibody and antigen titration, where two-fold steps produce the classic endpoint titre; of minimum inhibitory concentration testing, where two-fold antibiotic series are prescribed by the reference methods; and of every standard curve in an immunoassay or qPCR run. Once the series exists, converting plates back to a titre is handled by the CFU per mL calculator.

The formula, and the ratio that trips everyone up

The dilution factor of one step is the final volume divided by the sample volume: DF = (Vs + Vd) ÷ Vs. Transfer 1 mL into 9 mL and the final volume is 10 mL, so the factor is 10 — written 1:10, or 10⁻¹, or "a ten-fold dilution".

The mistake that survives into published protocols is reading 1:10 as one part sample to ten parts diluent. That combination gives a final volume of eleven parts and a factor of 11, not 10 — an error of 10% at every step, compounding to 77% after six steps (11⁶ ÷ 10⁶ = 1.77). Whenever the ratio matters, state the volumes rather than the ratio, and check that the numbers you write down sum to the final volume you intend.

Because each step multiplies the previous dilution, cumulative dilution after n identical steps is DFn, and the concentration in tube n is C0 ÷ DFn. Two properties follow. First, in logarithms the series is a straight line: log₁₀C falls by log₁₀DF at every step, which is why a ten-fold series drops exactly one log per tube and a two-fold series drops 0.301. Second, the arithmetic never depends on the tube volume you keep — only on the ratio you transfer. Doubling every volume in the protocol changes nothing about the dilutions, which is what makes a series scale freely from a 96-well plate to a rack of 50 mL tubes.

The formula also tells you what to do if the steps are not identical. Multiply the individual factors instead of raising one to a power: a 1:10 followed by a 1:5 followed by a 1:2 is a cumulative 1:100. This calculator assumes identical steps because that is what almost every protocol specifies; for a one-off mixed dilution use the C₁V₁ = C₂V₂ dilution calculator for each stage.

Worked example: a six-tube ten-fold series from 1 × 10⁶ cells/mL

You have a suspension at 1 × 10⁶ cells/mL and want plates in a countable range, so you build a decimal series: 1 mL transferred into 9 mL of diluent, six times.

  1. Dilution factor per step. (1 + 9) ÷ 1 = 10.
  2. Tube 1. 1 × 10⁶ ÷ 10 = 1 × 10⁵ per mL, a cumulative 1:10.
  3. Tube 2. 1 × 10⁶ ÷ 10² = 1 × 10⁴ per mL, cumulative 1:100.
  4. Tube 3. 1 × 10³ per mL, cumulative 1:1 000.
  5. Tube 4. 1 × 10² per mL, cumulative 1:10 000.
  6. Tube 5. 1 × 10¹ per mL, cumulative 1:100 000.
  7. Tube 6. 1 × 10⁶ ÷ 10⁶ = 1 cell per mL, cumulative 1:1 000 000.

The cumulative dilution is 10⁶ and log₁₀ of that is 6, so the series spans exactly six logs. Total diluent required is 6 × 9 mL = 54 mL, and each tube holds 10 mL after the transfer arrives but only 9 mL once you have taken the next transfer out of it.

Now pick the plating dilution. If you spread 0.1 mL, tube 4 delivers 100 per mL × 0.1 mL = 10 colonies, tube 3 delivers 100, and tube 2 delivers about 1 000. Tube 3 is the one that lands inside the 25–250 countable window, and tubes 2 and 4 bracket it — which is exactly why you plate three consecutive dilutions rather than betting on one.

Change the volumes and check the ratio still holds: 0.5 mL into 4.5 mL is also (0.5 + 4.5) ÷ 0.5 = 10, and consumes half the diluent. 100 µL into 900 µL is the same factor again in a microplate.

Choosing the step size and the number of steps

Match the step size to what you are measuring. Ten-fold steps cover ground quickly and are the standard for viable counts, where the titre may be unknown within several orders of magnitude. Two-fold steps give fine resolution and are prescribed for antimicrobial susceptibility testing and for antibody endpoint titres, where you need to distinguish adjacent concentrations rather than survey a wide range. Three- to five-fold steps are common for standard curves in immunoassays and dose-response work, where eight points should span the assay's dynamic range without wasting points outside it.

Choose the number of steps from the target, not from habit. If you need to fall from 10⁸ to 10² per mL, that is six logs, which is six ten-fold steps or twenty two-fold steps (2²⁰ = 1.05 × 10⁶). The two-fold route reaches the same place with far more pipetting and far more accumulated error, so use it only when you need the resolution.

Watch the volume you have left. Each tube gives up the transfer volume to the next tube, so a 10 mL tube in a 1-into-9 series has 9 mL available for assays — plenty. A 200 µL well in a two-fold plate series that transfers 100 µL leaves only 100 µL to read, which is why plate titrations are usually set up with the final volume in mind first and the transfer volume derived from it.

One structural point that catches people out: the last tube in the series never gives up a transfer, so it retains its full volume. If your protocol says to discard 100 µL from the final well to equalise volumes across the plate, that instruction exists for photometry, not for the dilution math — the concentration in that well is correct either way.

Common transfer volumes and the factors they give

Every row is DF = (sample + diluent) ÷ sample. The last two columns show how far the series reaches after the stated number of identical steps.
TransferInto diluentDF per stepCumulative after 6 stepsCumulative after 10 steps
1 mL1 mL2641 024
1 mL2 mL372959 049
1 mL3 mL44 0961 048 576
1 mL4 mL515 6259 765 625
1 mL9 mL101 000 00010¹⁰
0.5 mL4.5 mL101 000 00010¹⁰
100 µL900 µL101 000 00010¹⁰
1 mL10 mL111 771 5612.59 × 10¹⁰

The last row is the classic error: one part into ten parts of diluent is a 1:11 dilution, and after six steps it is 77% away from the 1:1 000 000 you intended.

What ruins a dilution series

  • Not changing the tip between steps. Carry-over on the outside of a tip is a fixed volume of concentrated material added to a progressively more dilute tube, so it dominates the far end of the series. Change tips at every transfer.
  • Failing to mix before the next transfer. Vortex or pipette-mix each tube thoroughly. A transfer taken from an unmixed tube samples whatever stratified at the top, and no arithmetic recovers from it.
  • Reading 1:10 as one part in ten parts of diluent. That is a factor of 11. Over six steps it compounds to a 77% error.
  • Letting cells settle in the stock. A suspension that sat on the bench for ten minutes is no longer at the concentration you measured. Mix the stock immediately before the first transfer.
  • Using a pipette below its calibrated range. Delivering 2 µL with a 200 µL pipette is far outside its accuracy specification. ISO 8655 sets tolerances by volume, and error grows sharply at the bottom of a pipette's range.
  • Assuming the diluent is inert. Some organisms lose viability rapidly in plain water or unbuffered saline. Use the diluent the method specifies, and plate within the holding time it allows.
  • Adding transfers you do not need. Every step adds error. Reach the target in the fewest steps the volumes allow.

How error builds along the series

Two kinds of error behave very differently down a series, and knowing which you have determines whether a long series is acceptable.

Random error — the ordinary scatter of pipetting — partially cancels, because a step that runs slightly high is as likely as one that runs slightly low. For independent errors of similar relative size at each step, the relative uncertainty of the cumulative dilution grows roughly with the square root of the number of steps. Ten steps carry about three times the relative uncertainty of one step, not ten times.

Systematic error does not cancel at all. A pipette that consistently delivers 2% low applies that bias at every step, so the cumulative dilution is out by 1.02n: 13% after six steps, 22% after ten. This is the argument for calibrating the pipette you use for a standard curve, and for using the same pipette for the whole series so that any residual bias is at least consistent.

There is also a floor you cannot pipette your way past. Carry-over on the outside of a tip, aerosol during vortexing, and adsorption of the analyte to tube walls all add material to tubes that should be nearly empty. In practice this sets a limit on how deep a manual series is meaningful, and it is why an extremely dilute tube should be verified against a direct measurement rather than trusted from the arithmetic alone. For cells, that verification is a plate count or a chamber count with the hemocytometer cell count calculator; for growth experiments, the two counts feed straight into the cell doubling time calculator.

Set the whole rack out before you start

Pre-dispense the diluent into every tube, label them with their cumulative dilution rather than their step number, and lay out one tip per step. Labelling tubes 10⁻¹ to 10⁻⁶ instead of 1 to 6 removes the most common transcription error at the plating bench, because the label then matches the number you will type into a calculation. The total diluent figure on this page tells you how much to pour before you begin — 54 mL for a six-step 1-into-9 series.

Key terms

Dilution factor
Final volume divided by sample volume for one step. A 1:10 dilution has a factor of 10 and reduces concentration to one tenth.
Cumulative (total) dilution
The product of every step's factor. For n identical steps it is DF raised to the power n.
Decimal dilution
A ten-fold step, the standard in microbiology because it moves the count by exactly one log₁₀ per tube.
Two-fold (doubling) dilution
A 1:2 step, used for antibody titres and MIC panels because it gives fine resolution across a narrow range.
Carry-over
Concentrated material transferred unintentionally, usually as a film on the outside of a tip. It sets the practical depth limit of a manual series.
Endpoint titre
The reciprocal of the highest dilution still giving a positive result — read directly off the cumulative dilution column of a two-fold series.

Frequently asked questions

Is 1:10 one part in ten or one part plus ten parts?

One part in ten total: 1 mL of sample plus 9 mL of diluent, giving 10 mL of final volume and a dilution factor of 10. One part sample plus ten parts diluent is 11 mL of final volume and a factor of 11. The distinction looks pedantic in a single step and is not pedantic at all in a series — six steps at 1:11 instead of 1:10 leaves you 77% away from the dilution you wrote in the notebook.

How many steps do I need to reach a target concentration?

Divide the log of the total dilution you need by the log of your step factor. Going from 10⁸ to 10² per mL is a 10⁶ total dilution; with ten-fold steps that is log(10⁶) ÷ log(10) = 6 steps, and with two-fold steps it is log(10⁶) ÷ log(2) = 19.9, so 20 steps. Enter your stock and step volumes above and read the cumulative dilution column of the table to find the tube that lands where you need it.

Can the dilution steps be different sizes?

Yes, and the cumulative dilution is then the product of the individual factors rather than a power. A 1:10 followed by a 1:5 gives 1:50. This calculator assumes identical steps because nearly every published protocol specifies them, which also makes the series easier to check and to label. For a mixed series, run each stage through a C₁V₁ = C₂V₂ dilution calculation and multiply the factors together.

Which tube should I plate for a viable count?

Plate at least three consecutive tubes and count whichever plate lands in your method's countable window — 25 to 250 colonies under FDA BAM, 30 to 300 under Standard Methods. With 0.1 mL spread from each tube, the plate you want carries roughly 250 to 2 500 organisms per mL in the tube. Plating three tubes costs three plates and guarantees a countable one; plating a single tube is a bet on already knowing the answer.

Why does my final concentration not match the calculation?

Ranked by how often it is the culprit: tips not changed between steps, tubes not mixed before the next transfer, the 1:10 versus 1:11 ratio error, a pipette used below its calibrated range, and settling in the stock between measurement and use. Carry-over and inadequate mixing both bias the far tubes upward, so a series that reads high at the dilute end almost always points at technique rather than at the arithmetic.

How do I make a two-fold dilution series in a 96-well plate?

Put 100 µL of diluent in every well of the row from column 2 onward, put 200 µL of sample in column 1, then transfer 100 µL from column 1 into column 2, mix, and carry 100 µL along the row. Each transfer gives (100 + 100) ÷ 100 = 2, so twelve wells span a 1:2 048 range. Discard the last 100 µL only if you need equal volumes for reading — the concentration in that well is correct whether or not you do.

What is a reasonable maximum number of steps?

Use the fewest steps that reach your target. Beyond about ten to twelve manual transfers, carry-over and accumulated systematic bias usually matter more than the nominal arithmetic, and a cumulative dilution past 10¹² is very hard to defend from a manual series alone. If you need to go deeper, dilute in two stages with a fresh tip, a fresh vessel and a verified intermediate, or verify the endpoint against a direct count.

Does this calculator work for chemical solutions as well as cells?

Yes. The dilution factor depends only on volumes, so the same series applies to a molar solution, a µg/mL protein stock or a cell suspension. Enter the stock concentration in whatever unit you use and read the concentration column in that unit. For molar solutions where you need mass and volume rather than a ratio, the molarity calculator and the ppm concentration calculator handle the unit conversions directly.

References

  • ISO 6887-1: Preparation of test samples, initial suspension and decimal dilutions for microbiological examination — International Organization for Standardization
  • ISO 8655: Piston-operated volumetric apparatus — accuracy and precision requirements — International Organization for Standardization
  • Bacteriological Analytical Manual, Chapter 3: Aerobic Plate CountU.S. Food and Drug Administration
  • Molecular Cloning: A Laboratory Manual, 4th ed. — Cold Spring Harbor Laboratory Press (Green & Sambrook)